CelCulture
24
IR Sensor Removal
IR sensor is located behind the blower in the top duct.
To remove the IR sensor:
1.
Unclip the sensor out of its place
2.
Disconnect the probe (1) from its cable (2)
TC and O
2
Sensor Removal
TC and O
2
Sensor are located behind the blower in the top duct
To remove the TC or O
2
sensor:
1.
Unscrew all 4 screws at the corners of the sensor assembly (1)
2.
Lower down the sensor assembly
3.
Disconnect the sensor connector (2)
4.
Replace the sensor assembly with the included blank plate (3)
5.
Screw the blank plate in place
Once the decontamination is done, reinstall all the sensors
1
2
1
2
3
Do not touch this part (slotted filter cap) of the sensor
Summary of Contents for CelCulture CO2