2.1 General Requirement
2.1.1 Location Requirements
The CelCulture CO
2
Incubator can be placed on the floor or working surface or floor stand. Do not
place equipment close to flammable materials or devices that produce excess heat.
Before moving or relocating the unit, make sure to:
o
Remove the water in the humidity to avoid spillage.
o
Transfer the sample cells to another incubator.
It is essential to ensure adequate air ventilation around the equipment.
When moving the equipment, do not lift using the door or door handle.
The spacing to the side wall should be at least 10 cm (3.9”).
The spacing at the back of the equipment should be at least 11 cm (4.3”).
2.1.2 Environmental Requirements
Indoor use
Altitudes of up to 2000 meters (6600’)
Up to 90% relative humidity – non condensing
Temperature between 18°C - 34°C (64°F - 93°F)
Note:
23°C - 27°C (73°F - 81°F) is the range of the best performance
Must be installed in room with sufficient ventilation. The room ventilation should be a technical
ventilation that is in accordance with the requirements of ZH 1/119 (Guidelines for laboratories) or
some other suited ventilation system with appropriate capacity.