13
CO
2
Incubator
3.2.1 Settings
3.2.1.1 Set Temperature
Your Incubator has an operating temperature range of a3°C to 60°C. The default temperature set
point is 37
o
C.
MENU
SETTINGS
SET TEMP
20 < X < 60
3.2.1.2 Set %CO2
Your Incubator has a range of 0% -20% CO₂. The default CO₂ set point is 5%.
MENU
SETTINGS
SET %CO2
0 < X < 20
3.2.1.3 Set %O2 (
for suppressed O
2
model
)
Your Incubator has a range of 1% -20.7% O₂. The default O₂ set point is 5%.
MENU
SETTINGS
SET %O2
1 < X < 20.7
3.2.2 Calibration
After the unit has stabilized (recommended to run overnight); temperature, and CO₂, O₂, and RH sensors can
all be calibrated to using a reference instruments. See
section 3.3.1 Unit Calibration
.
3.2.3 Admin
3.2.3.1 Over Temperature Alarm
The unit will display/sound the alarm when the temperature goes above the over temperature alarm value.
The default over temperature alarm is temperature set point + 1
o
C.
MENU
ADMIN
OVER TEMP ALARM
SP + 1 ≤ X ≤ 65
3.2.3.2 Low Temperature Alarm
The unit will display/sound the alarm when the temperature goes below the low temperature alarm value. The
default low temperature alarm is temperature set point - 1
o
C.
Note:
The low temperature alarm will only be activated once the unit reaches the temperature set point.
MENU
ADMIN
LOW TEMP ALARM
0 ≤ X ≤ SP - 1
3.2.3.3 High %CO2 Alarm
The unit will display/sound the alarm when the CO
2
level goes above the high %CO2 alarm value. The default
high %CO2 alarm is %CO2 set point + 1%.
MENU
ADMIN
HIGH %CO2 ALARM
SP + 1 ≤ X ≤ 21
3.2.3.4 Low %CO2 Alarm
The unit will display/sound the alarm when the CO
2
level goes below the low %CO2 alarm value. The default
low %CO2 alarm is %CO2 set point– 1%.
MENU
ADMIN
LOW %CO2 ALARM
0 ≤ X ≤ SP - 1
3.2.3.5 New CO2 Tank (
for unit without the optional gas backup switch system
)
The CO₂ tank depletion reminder automatically calculates how much CO₂ gas is left in the tank. It alerts the
user one week before the gas is depleted which gives the user some buffer time to order new tanks.
MENU
ADMIN
NEW CO2 TANK
XXXX