Polymerase Chain Reaction Cabinets
18
1.5.2
PCR-A
1.
Remove the black foam at top exhaust
2.
Unscrew the top metal mesh and remove
3.
Disconnect the blower wiring
4.
Unscrew the blower mounting house and take out
5.
Replace the blower and reverse above steps
6.
Re-certify the cabinet