Polymerase Chain Reaction Cabinets
Fig 2.3
Blower does not
function
Faulty motor speed
controller
•
Unplug the cabinet from main supply.
•
Open electrical panel and see Layout A/B at the end of
this section to locate the motor speed controller.
•
Add a jumper wire between the 2 cables going to speed
control (see Fig. 2.4, 2.5).
•
Plug the cabinet to the main supply.
•
If the fan operates properly, the speed controller is
defective and need to be replaced.
•
If the fan does not operate, the problem is not with the
speed control. Proceed to next step.
Fig 2.4
Fig 2.5