19
1.6
Fluorescent Lamp and UV Lamp replacement
1.6.1
Replacing the fluorescent lamp
1.
Before changing the fluorescent bulbs, ensure that the cabinet is powered down and disconnected
from the electrical supply
2.
Locate the bulbs.
3.
Remove the power clips at the ends of the bulbs by gently pulling whilst holding the bulb steady.
4.
Remove the bulbs from the mounting clips and replace with new ones.
5.
Replace the power clips on the ends of the new bulbs and ensure they are firmly seated.
6.
Reconnect the cabinet to the electrical supply and test the bulbs for proper operation.
1.6.2
Replacing the UV lamp
1.
Disconnect the electrical connections
2.
Take out the old UV lamp by rotating 90° counter-clockwise and pull downward.
3.
Install the new UV lamp by pushing it upward and rotate 90° clockwise.