1.3.1.2
HEPA/ULPA Filter Leak Test
The filter was challenged with PAO aerosol of fixed concentration. The aerosol was evenly distributed
throughout the supply (positive) cabinet plenum. An aerosol photometer was used to monitor aerosol
penetration downstream of the filter, and to scan for the presence of leaks.
Use calculated method, along with ATI TDA-4Blite aerosol generator, the upstream aerosol concentration
can be calculated using the following equation:
Aerosol concentration
=
13,500×𝑁
𝑇𝑜𝑡𝑎𝑙 𝐴𝑖𝑟𝑓𝑙𝑜𝑤 𝑐𝑓𝑚
, where N = Number of Laskin nozzle(s) opened
See Chart for Total Airflow below:
PCR cabinet model
2 feet
Total Airflow
177
Airflow Filter
Leaks detected in media
Yes / No
Leaks detected in gasket
Yes / No
Particle penetration
%
Acceptance:
Maximum particle penetration of 0.01% for filters; no filters leaks present.
Result:
Pass / Fail
Acceptance criteria / testing procedures have been adopted in accordance with the requirements of the:
a.
EN 1822 “High Efficiency Air Filters (HEPA and ULPA)”
b.
IEST-RP-CC034.3 “Recommended Practice for HEPA and ULPA Filter Leak Tests”
AS 1807.6 “ Filter Installation Integrity”
1.3.1.3
Particle Count Test
Use a particle counter with sampling rate of 0.028 m
3
(1 ft.
3
) of air per minute. Put the particle counter
sampling cone inside the work zone, following the grid below:
Actual PAO concentration:
g/liter
DOMINIQUE
DUTSCHER
SAS
Summary of Contents for Airstream PCR-3A
Page 6: ...D O M I N I Q U E D U T S C H E R S A S...
Page 13: ...D O M I N I Q U E D U T S C H E R S A S...
Page 34: ...D O M I N I Q U E D U T S C H E R S A S...
Page 57: ...44 Polymerase Chain Reaction Cabinets LAYOUT A D O M I N I Q U E D U T S C H E R S A S...
Page 58: ...45 LAYOUT B D O M I N I Q U E D U T S C H E R S A S...
Page 60: ...47 D O M I N I Q U E D U T S C H E R S A S...
Page 61: ...48 APPENDIX D O M I N I Q U E D U T S C H E R S A S...