Polymerase Chain Reaction Cabinets
34
Problem 6
—
Blank LCD
PROBLEM
POSSIBLE CAUSE
CORRECTIVE ACTION
Blank LCD
Connection problem
•
Unplug the cabinet from the main supply.
•
Open the top access cover and locate electrical panel
behind it.
•
Open electrical panel and see Layout A at the end of this
section to locate the main board inside electrical panel.
•
Check whether LCD FRC (Flat Ribbon Cable) has been
inserted properly into its socket on the main board (see Fig.
6.1).
Fig 6.1