Problem 9
—
Cannot turn on UV lamp
PROBLEM
POSSIBLE CAUSE
CORRECTIVE ACTION
Cannot turn ON UV
Lamp
Sash not in UV state
WARNING:
Before turning the UV lamp on, make sure that
the hinged window is closed. Also ensure that you are
protected.
For PCR:
•
Close the hinged window for UV mode position (fully
closed, see
Fig
. 9.1 below).
•
If UV MODE is not displayed on the LCD, please refer to
the Magnetic Switch troubleshooting section.
•
Switch on the UV lamp by pressing the UV button (see Fig.
9.2 below).
•
If UV MODE is displayed on LCD but UV is not on after
pressing UV button, please refer to possible cause 2-UV
magneticswitchis misaligned or faulty.
For SCR:
•
Close the hinged window for UV mode position (fully
closed, seeFigure 9-1 below).
•
Set the mechanical timer for the UV at the desired value.
•
If UV timer is on but UV lamp is not, please refer to possible
cause 2-UV magnetic switch is misaligned or faulty.
Fig 9.1
Fig 9.2