Blower does not
function
Auto-thermal cut-off
•
Check the Fan for overheating – The motor has an in-
built thermal cut off.
•
Wait 60 minutes with the FAN turned off and then try to
restart.
•
If the FAN restarts determine why there is excessive
heat in the cabinet.
•
If the FAN does not restart proceed to the next step.
Motor failure
•
Unplug the cabinet from the main supply.
•
Check motor condition. If physical damage is present, like
fan blades bent, cracked or chipped off, replace blower
(see Fig. 2.9).
•
To replace the blower, please follow instruction which is
available in this manual under Service Section.
Fig 2.9
DOMINIQUE
DUTSCHER
SAS