9 Control Terminal T
ergoselect 4 / 5
77
HR Monitoring
These parameters are required for the “Interval HR” training
mode.
If the heart rate exceeds the "Heart Rate" value (see
figure 9 – 37, Type [ Interval HR ] on page 59) by the
"Threshold" value for the duration of the "Delay", the load
for the next interval will be reduced:
• by a relative value [%] or
• by an absolute value [W]
Delay
Threshold
Type
Reduction
5 sec
3 1/min
Relative
4 %
HR monitoring
Figure 9 – 85: Defining the heart-rate monitoring parameters
Copy HR Belt
This function can be used to transfer the ID numbers of all
16 stored heart rate belts to another ergometer.
To do this, insert a chip card and then press the
Export to
CPC
key. All ID numbers will be copied from the memory of
the ergometer to the chip card. First, however, check that
there are no training data on the chip card, as they will be
overwritten and become useless.
Then remove the chip card and insert it in the next ergometer.
To import the ID number, press the
Import from CPC
key.
All the ID numbers on the chip card will now be transferred
to the ergometer memory. The ID numbers can be imported
at any number of devices.
Export to CPC
Import from CPC
Copy HR belt
Please insert card!
Figure 9 – 86: Copy HR belt
Quick Start
When the
Quick Start
function is activated, a manual training
session will start automatically only after the pedal speed
exceeds 30 rpm.
Off
On
Quick Start
Figure 9 – 87: Activating the Quick Start mode