12 General Product Information
82
ergoselect 4 / 5
12 General Product Information
12.1 Checks Before Each Use
Before each use, visually inspect the device for signs of
mechanical damage. If you detect damage or impaired
functions which may result in a hazard to the patient or
the operator, the device must be repaired before it can be
used again.
The parts of the device visually inspected shall be free of
cracks or signs of damage, abrasion or wear.
The visual inspection includes the following parts of the
device:
• housing
• equipment cables and power cord
• membrane keyboard
• electrode insulation and connectors
• input / output ports, plug-in connections
• connecting cables
In addition to the visual inspection the device should be
switched on and the proper functioning of the control
panel should be verified. This will ensure
• the proper functioning of the device
• the proper functioning of the display.
Warning
It is not permitted to use defective devices or
damaged cables. They shall be replaced immedi
ately.
12.2 Technical Safety Inspections and
Inspections of the Measuring System
The technical safety inspections and the inspections of
the measuring system must be completed every two years
according to the rules of the art by a Service Engineer
authorized by ergoline.
Similarly, the automatic sphygmomanometer in the control
terminal must be checked and, if necessary, calibrated by
an authorized specialist every two years to fulfill legal
requirements.
The date of the next inspection is indicated on the inspection
sticker attached next to the type plate on the ergometer.