2 • Safety Features
20
/ 54
8
0113-90
1
08
.1
1
•
If a button on the front panel has short-circuited due to an error or was pressed
when the power switch was switched on, this error will be indicated acoustically
and by an error number and message after switch-on of the power switch.
•
If a contact of the footswitch has short-circuited due to an error, or a pedal is
jammed or a pedal was pressed while the power switch was switched on, this
error will be indicated acoustically and by an error number and message.
The relevant error message on the display of the VIO tells you how to remedy the
error.