8 • Installation
36
/ 54
8
0113-90
1
08
.1
1
CAUTION
Insufficient acclimatization time, unsuitable temperature
during acclimatization
If the device was stored or transported below or above a certain
temperature, it will take a certain time and temperature to acclima-
tize.
If you do not observe the rules, the device can sustain damage and
fail.
¨
Acclimatize the device according to the rules in the Technical Da-
ta.
CAUTION
Overheating of the device due to poor ventilation
If ventilation is poor, the device can overheat, sustain damage, and
fail.
¨
Install the device in such a way that there is an unobstructed cir-
culation of air around the housing. Installation in confined wall
recesses is prohibited.
CAUTION
Penetration of liquid into the device
The housing is not absolutely watertight. If liquid penetrates, the
device can sustain damage and fail.
¨
Make sure no liquid can penetrate the device.
¨
Do not place vessels containing liquids on top of the device.
Electrical installation
WARNING
Defective grounded power outlet, power supply network
without proper grounding, inferior-quality power cord,
incorrect line voltage, multiple power outlets, extension
cords
Risk of electric shock and other injuries to the patient and medical
personnel! Risk of damage to property.
¨
Connect the unit / the equipment cart to a properly installed
grounded power outlet.
¨
Only connect the unit to a power supply network with proper
grounding.
¨
Only use the ERBE power cord or an equivalent power cord for
this purpose. The power cord must bear the applicable national
test symbol.
¨
Check the power cord for damage. You must not use a damaged
power cord.
¨
The supply voltage must match the voltage specified on the
unit's rating plate.
¨
Do not use multiple power outlets.
¨
Do not use extension cords.
Summary of Contents for VIO 300 D
Page 2: ......
Page 3: ...VIO 300 D Configured for da Vinci User Manual...
Page 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Page 22: ...3 Accessories 22 54 80113 901 08 11...
Page 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Page 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Page 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Page 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Page 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...