1 • Safety Instructions
10
/ 54
8
0113-90
1
08
.1
1
CAUTION
Overheating of the device due to poor ventilation
If ventilation is poor, the device can overheat, sustain damage, and
fail.
¨
Install the device in such a way that there is an unobstructed cir-
culation of air around the housing. Installation in confined wall
recesses is prohibited.
CAUTION
Penetration of liquid into the device
The housing is not absolutely watertight. If liquid penetrates, the
device can sustain damage and fail.
¨
Make sure no liquid can penetrate the device.
¨
Do not place vessels containing liquids on top of the device.
Electric shock
WARNING
Defective grounded power outlet, power supply network
without proper grounding, inferior-quality power cord,
incorrect line voltage, multiple power outlets, extension
cords
Risk of electric shock and other injuries to the patient and medical
personnel! Risk of damage to property.
¨
Connect the unit / the equipment cart to a properly installed
grounded power outlet.
¨
Only connect the unit to a power supply network with proper
grounding.
¨
Only use the ERBE power cord or an equivalent power cord for
this purpose. The power cord must bear the applicable national
test symbol.
¨
Check the power cord for damage. You must not use a damaged
power cord.
¨
The supply voltage must match the voltage specified on the
unit's rating plate.
¨
Do not use multiple power outlets.
¨
Do not use extension cords.
WARNING
Incorrect line fuse, defective device
Risk of electric shock to the patient and medical personnel! Risk of
damage to property.
¨
Blown line fuses may only be replaced by a competent techni-
cian. Only replacement fuses that have the same rating as the
one specified on the unit’s rating plate may be used.
¨
When a fuse has been changed, the function of the unit must be
verified. If the unit does not function properly or if there are any
concerns, please contact ERBE.
Summary of Contents for VIO 300 D
Page 2: ......
Page 3: ...VIO 300 D Configured for da Vinci User Manual...
Page 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Page 22: ...3 Accessories 22 54 80113 901 08 11...
Page 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Page 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Page 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Page 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Page 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...