Maintenance
Modifications and repairs
Modifications and repairs must not impair the safety of the equipment or equipment
cart and accessories for the patient, user and the environment. This condition is met
when changes to the structural and functional characteristics are not detrimental to
safety.
Authorized persons
Modifications and repairs may only be undertaken by ERBE or by persons expressly
authorized by ERBE. ERBE accepts no liability if modifications and repairs to the unit
or accessories are made by unauthorized persons. This will also invalidate the war-
ranty.
Technical safety checks
The technical safety checks determine whether the safety and operational readiness
of the unit or the equipment cart and accessories conform to a defined technical re-
quired status. Technical safety checks must be performed at least once a year.
What technical safety checks
must be performed?
For this device the following technical safety checks have been stipulated:
•
Checking of labels and User Manual
•
Visual inspection of unit and accessories for damage
•
Testing the grounded conductor in accordance with IEC 60601-1 Section 18
•
Testing the leakage current in accordance with IEC 60601-1 Section 19
•
Functional testing of all the unit's operating and control elements
•
Testing the monitoring equipment
•
Measurement of DC resistance
•
Testing the footswitch activation
•
Testing the spark monitor
•
Measurement of power output in COAG mode
The results of the safety checks must be documented.
If during the safety checks any defects are found which might endanger patients,
staff or third parties, the device may not be operated until the defects have been rem-
edied by competent service technicians.
Disposal
Your product bears a crossed-out garbage can icon (see image). Meaning: In all EU
countries this product must be disposed of separately in accordance with the nation-
al laws implementing EU Directive 2002/96/EC of January 27, 2003, WEEE.
In non-EU countries the local regulations must be observed.
Summary of Contents for VIO 300 D
Page 2: ......
Page 3: ...VIO 300 D Configured for da Vinci User Manual...
Page 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Page 22: ...3 Accessories 22 54 80113 901 08 11...
Page 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Page 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Page 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Page 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Page 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...