49
/ 54
12 • Information on electromagnetic compatibility (EMC)
8
0113-90
1
08
.1
1
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location
in a typical commercial or hos-
pital environment.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity
Guidance and manufacturer's declaration - electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the
equipment should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
- guidance
Portable and mobile HF commu-
nications equipment should be
used no closer to any part of the
equipment, including cables,
than the recommended separa-
tion distance. The separation dis-
tance is calculated from various
equations depending on the fre-
quency of the portable and
mobile HF communications
equipment:
Recommended separation
distance
Conducted HF IEC 61000-
4-6
3 V
rms
150 kHz to 80 MHz
3 V
rms
Equation 1) d=1.2 P
1/2
Radiated HF IEC 61000-4-
3
3 V/m
80 MHz to 800 MHz
3 V/m
Equation 2) d=1.2 P
1/2
3 V/m
800 MHz to 2.5 GHz
3 V/m
Equation 3) d=2.3P
1/2
Summary of Contents for VIO 300 D
Page 2: ......
Page 3: ...VIO 300 D Configured for da Vinci User Manual...
Page 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Page 22: ...3 Accessories 22 54 80113 901 08 11...
Page 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Page 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Page 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Page 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Page 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...