ELite 3, User Manual
Page 53/72
8.
Diagnostics
Diagnostics menu allows access to system information and hardware check-up.
Maintenance
Diagnostics
Press HOME to go to Main menu.
Press BACK to go back to previous menu.
8.1.1.
Device Information
Device information shows system hardware and software setup.
Maintenance
Diagnostics / Device Information
The various system parameters can be seen
on the screen.
Press EXIT to go back to the previous menu.
8.1.2.
Self Test
Self test is a procedure to verify proper operation of essential components of the
instrument. Self test should be performed:
•
At installation.
•
After replacing any component.
•
After extended time out of use.
During self test, the analyzer checks system components, and displays the results.
On the right side of the results screen, the SW displays if the tested parameter falls
into the desired range:
•
If yes, a
sign is displayed at the end of the line, or
•
if it is out of range: a
sign appears.