ELite 3, User Manual
Page 48/72
7.2.1.
Calibration with factors
Factor based calibration allows adjustment of primary parameters with a factor.
Maintenance
Calibration / Factors
Press white data field to modify calibration
factor. A numeric input screen will show up
so that you can enter values.
All values must be in the 0.8…1.2 range.
Press Accept to proceed with new
settings, or Cancel to keep values
unchanged.
7.2.2.
Calibration by measurement
The analyzer can run calibrator, and perform calculation of factors automatically.
User can decide the number of measurements to use for calibration.
Maintenance
Calibration / Measurement
Prior
to
starting
the
calibration
measurements, you have to define some
basic parameters for the upcoming
measurements.
As the next step, target values from the assay sheet of the calibrator or control must
be entered before measurement.
Maintenance
Calibration / Measurement
Prior
to
starting
the
calibration
measurements, you have to define some
basic parameters for the upcoming
measurements.