ELite 3, User Manual
Page 49/72
Target values for calibrated parameters can be set within the following ranges:
Parameter
Low limit
High limit
WBC
1.0
30.0
RBC
1.00
8.00
HGB g/l
30
300
MCV
50
120
RDW CV
10
50
PLT
30
800
MPV
5
15
PDW CV
5
50
HCT
0.1
0.6
PCT
0
2
Table 7. Calibration target ranges
When all parameters are set, press Accept key.
The display shows Calibration measurement at top.
Maintenance
Calibration / Measurement
Insert the sample tube to the sample door
and press the START button.
Exit will abort the operation.
Calibration runs are saved automatically. If you find that a result should not be used,
use the Discard button to delete the measurement so that it is not used for
calibration.
Maintenance
Calibration / Measurement / Result
Result will display the average of each
parameter of accepted measurements
compared to the target value and the
calibration factor calculated.
Accept saves new factors and aborts
calibration
Back will return to the calibration
measurement screen so that you can
measure more samples for calibration.
You can compare target and measured values, observe CV and see how the
calibration factor would change.
Maintenance