3.11.15.8
Configuration step by step
Starting point
In menu
Function Generator
find the PV functions. Select here group
PV EN50530 (PS)
.
Step 1: Technology selection
The extended PV function requires to select the panel technology of
the solar panel which is going to be simulated. In case
cSI
or
Thin Film
don’t match your requirements or you are not sure about their technol-
ogy parameters, select
Manual
.
When selecting
Thin film
or
cSI
the configuration continues with
Step
2
.
Step 1-1: Adjust technology parameters
If technology
Manual
was selected in the previous screen, all dis-
played technology parameters can be adjusted by tapping on them
and entering the desired value. It’s recommended to adjust these val-
ues very carefully, because wrong settings can result in a PV curve
which doesn’t work as expected.
When resetting the device, these values are reset to defaults which are
the same as with technology
cSI
. Also see the overview in
3.11.15.3. It
means they don’t necessarily have to be adjusted. If any of the other
technologies was selected, this screen would be skipped and these
parameters set to the defined values.
Step 2: Input mode and solar panel base parameters
Selecting the input mode between
MPP
and
ULIK
determines what
pair of parameters has to be set in the configuration and also later in
the simulation. When setting pair Uoc/Isc, the other two pair is calcu-
lated with factors and automatically set.
The open circuit voltage (
Uoc
) and the short-circuit current (
Isc
) are
upper limits which are usually read from the data sheet of a solar panel
and entered here for the simulation. Two parameters each are con-
nected via the fill factors:
U
MPP
= U
OC
× FFu / I
MPP
= I
SC
× FFi
Step 3: Select simulation mode
For a description of the available simulation modes see
When selecting
E/T
or
U/I
the configuration continues with
Step 4
,
otherwise an additional step is required