If your material tray has the serial number:
1. Tap
[With S/N]
.
2. Find the serial number on the front of your material tray and enter it to the field on the screen that appears.
3. Tap
[OK]
.
→ The
following screen appears.
IMPORTANT
Make sure your material tray
does not contain
material in it. Material residue in the material tray
during the calibration will give a false reading to the sensors.
4. Tap
[OK]
.
→ The platform starts moving to its Dry position.
→ Once the platform finds its Dry position, it starts moving to Zero
position.
→ Dry position calibration has been saved.
→ The following screen appears.
If your material tray does not have the serial number:
1. Tap
[Without S/N]
.