Safety Hazards
Warning hot Surface
During operation, the temperature of the probe filter head and of all parts exposed to flue gas is 150 °C –
600 °C (302 °F-932 °F). Direct contact with the hot parts for dismantling or maintenance will cause se-
vere burns!
The probe may only be removed with heat-insulated gloves. Before removing the probe, always switch off
the supply voltage to the electronic system. After removal, store the probe in a safe, protected place and
wait until it has cooled downs below 35 °C (95 °F).
1.5
Disruption of the Process
The analyzer system has to be kept in operation also in the event of the process being disrupted or if the plant is
powered off temporarily (e.g. at night or during the weekend). Frequently cooling down and heating up of the probe
results in thermal stress of the hot probe parts (heater, thermocouple and sensor) and reduces their product life.
ENOTEC will not accept any responsibility for resultant damage.
1.6
Storage instructions
ENOTEC equipment and spares are to be stored in a dry and ventilated environment at temperatures between
-40 °C to +80 °C (-40 °F to 176 °F). Paint fumes, silicone sprays, etc. must be avoided in the storage environment.