Status Messages
Installation and Operation Manual - COMTEC
®
6000
46
Doc.-ID: COM6000_11022020_EN
6.2
Alarm Messages
Error
Message
Relay
contact
Description
Reference air
flow too low
Possible Cause:
reference air flow set incorrectly, instrument air supply is
inadequate, reference air pump is faulty.
Reference air
flow too high
Possible Cause:
reference air flow is set incorrectly
Limit alarm 1
O
2
limit
alarm 1,
open
Possible Cause:
Over or under range of the O
2
limit alarm.
Limit alarm 2
CO
e
limit
alarm 2,
open
Possible Cause:
Over or under range of the CO
e
limit alarm.
Electronic
temp. too low
Possible Cause:
The ambient temperature of the transmitter is lower than the
specified lower limit. The specified measurement tolerances are no longer
guarateed.
Electronic
temp. too high
Possible Cause:
The ambient temperature of the transmitter is higher than the
specified upper limit. The specified measurement tolerances are no longer
guarateed.
Clock battery
low
The alarm can only be reset by the user after replacing the clock battery (Lithium
type 2032). As long as the system is connected to AC power, the alarm has no
impact. Only after restarting system will the time / date be incorrect. A possible timed
automatic calibration can no longer work correctly.
Unlisted alarm messages: Other messages cannot be remedied by the customer.
Please contact an ENOTEC service point
6.3
Service Messages
Maintenance
message
Relay
contact
Description
Measured
value(s) held
Service,
closed
When the measured values held is set to on, the determined O
2
mA output remains
saved for a duration before a calibration.