TOCII CA72TOC
Commissioning
Hauser
47
Settings for the analog outputs
You must enter the following parameters in the Basic Data menu:
PROGRAMMING/SETTING/BASIC DATA:
Settings for the alarm limits
You must enter the following parameters in the ALARM LIMITS menu:
PROGRAMMING/SETTING/ALARM LIMITS:
All the alarm limits affect the same signal output II (binary out II). A limit value alarm is also retained
following channel switchover until the limit value for the channel in question is undershot.
Parameter
Unit
Description
SCALE AO CH1
Scale end value of the analog output of channel 1.
SCALE AO CH2
Scale end value of the analog output of channel 2.
Parameter
Unit
Description
HI ALARM LIMIT CH1
mg/l
Use this function to specify the alarm value for when the value for
channel 1 is overshot.
LO ALARM LIMIT CH1
mg/l
Use this function to specify the alarm value for when the value for
channel 1 is undershot.
HI ALARM LIMIT CH2
mg/l
Use this function to specify the alarm value for when the value for
channel 2 is overshot.
LO ALARM LIMIT CH2
mg/l
Use this function to specify the alarm value for when the value for
channel 2 is undershot.