TOCII CA72TOC
Commissioning
Hauser
43
3
Analog inputs
• Current CO
2
measured value
• Furnace temperature
• Pressure level in gas circuit
• pH value in stripping vessel
Switch inputs
Switching status of the switch inputs
Ix = 0 = OFF
Ix > 0 = ON
PE = Peltier element
WA = Water pressure alarm
SB = Standby
LM = Leak detector
O2 = Gas pressure sensor
MEASUREMENT ON/OFF = Indicates
whether measuring operation is still
active in the background or not.
If you test the inputs when background
operation is active (e.g.: jumpering the
leak detector), the appropriate error
response measures are triggered by the
program.
Terminate the measurement operation
(Prog/Test/Measurement off) before
physically testing the input signals.
Step
Display
Description
=
=
=
=
xxx.xx
xxx
xxx
x.xx
IR
t
P
pH
ppm
°C
mbar
f =
f =
f =
f =
xxxx Hz
xxxx Hz
xxxx Hz
xxxx Hz
I1 =
I2 =
I3 =
I4 =
I5 =
I6 =
I7 =
I8 =
I9
I10
I11
I12
I13
I14
I15
I16
=
=
=
=
=
=
=
=
1
2
4
0
0
0
64
128
0
0
0
0
0
0
0
0
PE
PE
WA
SB
LM
O2
DIGITAL INPUTS
MEASUREMENT ON/OFF