Maintenance
TOCII CA72TOC
116
Hauser
Step
Display
Description
1
2
1.
Release the couplings (items 1 and
3).
2.
Replace the gas filter (item 2).
Take the direction of flow into
account.
3.
First secure the filter in the coupling
(item 3) and then connect the filter
to the acid filter with the coupling
(item 1). Make sure that the cone is
properly positioned on the filter.
4.
Tighten the two couplings.
Operation starts (initially without a
measured value).
PUMPS
CALIBRATION
S E R V I C E
CLEANING
FILTERS
REPLACE ACID FILTER
REPLACE GAS FILTERS
REPLACE O2 FILTERS
REPLACE HEATED FILTER
REPLACE GAS FILTERS
[E] TO CONFIRM
MAINTENANCE COMPLETED ?