TOCII CA72TOC
Commissioning
Hauser
45
6.3.13
Menu TEST - TEST FREQUENCY I/O
a0012616-en
Fig. 25:
"Test" menu
Test the inputs and outputs to test the function of the I/O card.
6.3.14
Menu TEST - TEST COM2
a0012617-en
Fig. 26:
"Test" menu
Displays the transmission data for the computer interface RS 232.
The menu item allows users to test the transmission of data with an external terminal.
Once the data connection has been established, the line "abcdef-ABCDEF-01234567890" can be
sent to the terminal by pressing a key, e.g. the [0] key.
Keystrokes at the external terminal are shown on the display.
Press the [E] key to exit the menu item.
Output
Description
Switch status OFF
Switch status ON
SA1
Changeover between standard and sample
Sample
Standard solution
SA2
Flushing valve for power flushing
Strip chamber flushing
switched off
Strip chamber flushing
switched on
SA3
O
2
supply, tube furnace controller, Peltier
cooler controller, membrane compressor
Consumer load switched off
Switch status during
measuring operation
SA4
Changeover between standard 1 and
standard 2
Standard 1
Standard 2
SA5
Screen flushing valve
Screen flushing valve off
Screen flushing valve on
SA6
Optional: Changeover between channel 1
and channel 2
Channel 1
Channel 2
SA7
Carrier gas flushing valve
Carrier gas flushing off
Carrier gas flushing on
SA8
Dosing valve
Dosing valve open
Dosing valve closed
SETTING
LISTS
P R O G R A M M I N G
TEST
TEST OF INPUTS
TEST OF OUTPUTS
TEST FREQUENCY I/O
TEST COM2
MEASUREMENT OFF
SETTING
LISTS
P R O G R A M M I N G
TEST
TEST OF INPUTS
TEST OF OUTPUTS
TEST FREQUENCY I/O
TEST COM2
MEASUREMENT OFF