TOCII CA72TOC
Maintenance
Hauser
65
3
1.
Set the sample valve to manual
sampling.
2.
To empty the pump hose and the
strip chamber, place a collecting
vessel under the hose connection for
manual sampling.
3.
Press the [E] key.
4
For version "with predilution":
1.
First open the hose cassettes for
pump 1.
2.
Now open the hose cassette for pump
4.
3.
Drain the strip chamber.
4.
Press the [E] key.
For version "without predilution":
1.
Open the hose cassette for pump 1.
2.
Drain the strip chamber.
3.
Press the [E] key.
Step
Display
Description
REPLACE HOSE P1/4
[E] TO CONFIRM
SWITCH VALVE TO MANUAL SAMPLING,
PUT COLLECTING VESSEL BELOW
REPLACE HOSE P1/4
[E] TO CONFIRM
OPEN HOSE CASSETTE(S) P1
OPEN HOSE CASSETTE P4
UNTIL STRIP CHAMBER IS DRAINED
REPLACE HOSE P1/4
[E] TO CONFIRM
OPEN HOSE CASSETTE(S) P1
UNTIL STRIP CHAMBER IS DRAINED