NS 5.14 Firmware
2
–
36
4900002262 rev. B 3-23-21
The value of the
RATA Multiplier
parameter,
S
, is determined by
,
where
C
1
is the certified concentration of standard No. 1,
C
2
is the certified
concentration of standard No. 2,
A
1
is the measured concentration (analyzer
reading) of standard No. 1 without any RATA adjustment, and
A
2
is the
measured concentration (analyzer reading) of standard No. 2 without any RATA
adjustment.
The
RATA Offset
parameter,
O
, is determined by
,
where
S
can be
1
when a sample from only one concentration standard is
available.
For a non-zero single concentration standard
RATA Multiplier
parameter,
S
is
determined by
where
O
can be zero, if desired.
After an automatic, semi-automatic, manual
Mode 7
or manual
Mode 8
validation completes, the analyzer automatically calculates new
RATA
Multiplier
and
RATA Offset
values. These values are based on the number of
validations that the analyzer is configured to accept and the type of validation
used.
To perform this calculation, the most recent validation result(s) will be used.
These calculations are based on the average measurement values during the
validation, which can be viewed from the
Mode 9
display. These values are not
constrained by the measurement range of the analyzer, which will ensure
accurate
RATA Multiplier
and
RATA Offset
values are calculated.
To perform the calculation
If RATA is enabled, the current
RATA Multiplier
and
RATA Offset
values must
be removed from the measured Validation 1 and/or Validation 2 concentration
values. Depending on the number of validations, or the validation type, one of
the following scenarios occur:
•
If using a non-permeation single validation system with a zero gas as
the standard, then a new
RATA Offset
is calculated leaving the
RATA Multiplier
at its previous value.
•
If using a non-permeation single validation system with a non-zero
gas as the standard, then a new
RATA Multiplier
is calculated
leaving the
RATA Offset
at its previous value.
S
C
2
C
1
–
A
2
A
1
–
--------------------
=
O
C
1
S A
1
–
=
S = (C
1
- O) / A
1
Summary of Contents for SpectraSensors SS2000e
Page 2: ......
Page 4: ......
Page 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 113: ......
Page 114: ...www spectrasensors com contact 4900002262 Rev B...