NS 5.14 Firmware
2
–
40
4900002262 rev. B 3-23-21
Validating the Analyzer
Validation of the analyzer using an appropriate gas standard is automatically
conducted at each
Val Interval
period at the
Val Start Time
, semi-
automatically when initiated via the
Validation DI
or the
Start Validation
parameter, or manually by accessing
Mode 7
or
Mode 8
.
To validate automatically
1.
Verify that the validation gas source(s) has been properly installed
and/or connected.
2.
If necessary, set the current time, date, desired validation hour and
validation interval (see
“To change parameters in Mode 2”
).
3.
Set the
Daily Validation
parameter to
1
. The analyzer should run
a validation cycle for the number of days and hour set by the
Val
Interval
and
Val Start Time
parameters.
To validate semi‐automatically
1.
Verify that the validation gas source(s) has been properly installed
and/or connected.
2.
Initiate a validation cycle by closing the contacts connected to the
Validation DI
input or by setting the
Start Validation
parameter
to
1
.
To validate manually
1.
Verify that the validation gas source(s) has been properly installed
and/or connected.
2.
Initiate validation measurements by pressing the
#
key
followed by
the
7
key (
Mode 7
) or the
#
key followed by the
8
key (
Mode 8
)
for a dual validation system.
3.
Once validation measurements are complete, press the
#
key
followed by the
1
key (
Mode 1
) to return to Normal Mode and stop
the validation.
Once a validation measurement has been completed, the appropriate relay (Val
#1 Active or Val #2 Active) will be activated. If the concentration measured
does not fall within the allowable limits (refer to
Validation Fail Alarm
relay will activate. Once activated, the
Validation Fail Alarm
must be manually canceled (refer to
Summary of Contents for SpectraSensors SS2000e
Page 2: ......
Page 4: ......
Page 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 113: ......
Page 114: ...www spectrasensors com contact 4900002262 Rev B...