NS 5.14 Firmware
2
–
16
4900002262 rev. B 3-23-21
2 Way Com Port
The
2 Way Com Port
parameter sets the port that allows two-way
communications, including Modbus and the diagnostic protocol. Enter
0
to turn
off two-way communications,
1
for the customer port,
2
for the service port
and
3
for the Ethernet port, if applicable.
The customer port baud rate is set from the Baud Rate parameter with 8 data
bits, 1 stop bit and no parity. The service port baud rate is 115200 with 8 data
bits, 1 stop bit and no parity. If the Ethernet port is available, refer to
“Configuring the Built-in Ethernet Port”
4-20 mA Alarm Action
The
4-20 mA Alarm Action
determines the current loop state upon an alarm
condition. Enter
0
for no action,
1
for the current loop to assume a low state
upon an alarm condition,
2
for the current loop to assume a high state upon an
alarm condition, or
3
for the current loop to track and hold the current state
upon an alarm condition.
4-20 mA Val Action
The
4-20 mA Val Action
parameter sets the operation mode of the 4-20 mA
current loop during validation cycles. Enter
0
for the current loop to track and
hold the last process measurement or
1
for the current loop to continue to
output the analyzer measurements during the validation cycle.
<SET PARAMETER MODE>
2 Way Com Port
0
0:Off1:Cus2:Ser3:Eth
<SET PARAMETER MODE>
4-20mA Alarm Action
0
0:None 1:L 2:H 3:T&H
<SET PARAMETER MODE>
4-20 mA Val Action
0
0:Hold 1:Measure
Summary of Contents for SpectraSensors SS2000e
Page 2: ......
Page 4: ......
Page 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 113: ......
Page 114: ...www spectrasensors com contact 4900002262 Rev B...