NS 5.14 Firmware
2
–
22
4900002262 rev. B 3-23-21
DO Alarm Setup
The
DO Alarm Setup
parameter sets the functionality of the
Assignable
Alarm
. Add the hexadecimal values according to Table 2–3 below for each fault
chosen to trigger the
Assignable Alarm
. Convert the resulting hexadecimal
value to a decimal value and enter the number for normally deactivated relay
functionality. Add ‘1’ to the resulting decimal value to switch to normally
activated functionality.
In the example above, the hexadecimal value of 0002000 converts to decimal
value of 8192, which when entered results in a normally deactivated relay
triggered by the
Concentra High Alarm
. Entering a value of 8193 would
result in a normally activated relay (fail safe) triggered by the
Concentra High
Alarm
.
The ISO method is only valid for sample pressures greater than
10000 mBar (10 bar), and therefore should not be used for
samples with pipeline pressures less than 10000 mbar. The
analyzer firmware assumes that any pipeline pressure less than
10 bar is the same as 10 bar, so the measured Dew Point value
will not change for sample pressures from 0 to 10 bar. For Natural
Gas sample pressures in the 0 to 10000 mbar range,
SpectraSensors recommends users select ASTM1 or ASTM2
methods. For air or nitrogen samples, SpectraSensors
recommends the Arden Buck method (Dew Point Method = 3).
Table 2–3 Assignable Alarm
functionality
Bit
Hex Value
Functionality
0
0000001
Normally closed or Power Fail (always on).
1
0000002
Any alarm active (general fault).
2
0000004
Laser power is too low.
3
0000008
Laser power is too high.
<SET PARAMETER MODE>
Dew Point Method
0
0:ISO1:AS12:AS23:AB
<SET PARAMETER MODE>
DO Alarm Setup
8192
Enter decimal value
Summary of Contents for SpectraSensors SS2000e
Page 2: ......
Page 4: ......
Page 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 113: ......
Page 114: ...www spectrasensors com contact 4900002262 Rev B...