NS 5.14 Firmware
2
–
28
4900002262 rev. B 3-23-21
2
), and then pressing the
*
key (without entering a password) to display the
first parameter.
Peak Tracking
The
Peak Tracking
function is a software utility that periodically adjusts the
laser current to keep the absorption peak of the measured component at a
known location. There are three choices:
0
for no peak tracking,
1
for peak
tracking (default), and
2
to reset the peak to its factory default setting.
Selecting
2
will return the current analyzer midpoint to the factory default
midpoint, and then automatically revert the parameter value to its setting
before the reset was initiated. In most cases, the peak tracking should be set
to
1
for on.
Pipeline Pressure
The
Pipeline Pressure
parameter sets the pipeline pressure (in mbar) in the
current dew point calculation or, if enabled, displays the current pipeline
pressure input through the
AI Pressure Input
. Enter the required value.
Pressure Unit
The
Pressure Unit
parameter designates the display units for the measured
absolute pressure in the cell. There are four choices:
0
for millibar,
1
for Torr,
2
for kPa, and
3
for PSIA.
<SET PARAMETER MODE>
Operator Password
0
0:No p/w >0:p/w
<SET PARAMETER MODE>
Peak Tracking
1
0:Off 1:On 2:Rst
<SET PARAMETER MODE>
Pipeline Pressure
0.00000
Enter a value (mb)
<SET PARAMETER MODE>
Pressure Unit
0
0:mb1:Torr2:kPa3:psi
Summary of Contents for SpectraSensors SS2000e
Page 2: ......
Page 4: ......
Page 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 113: ......
Page 114: ...www spectrasensors com contact 4900002262 Rev B...