Operating the Analyzer
Operator’s Manual
2
–
13
Table 2-1
Typical values for parameter setpoints (Continued)
Parameter
Setting
Function
AO 4 mA Value
-1.E+06 to 1.E+06
Default 0.0
Sets ppmv value corresponding
to 4 mA current loop output.
AO 20 mA Value
-1.E+06 to 1.E+06
Default 100.0
Sets ppmv value corresponding
to 20 mA current loop output.
Calculate Dew Point
0, 1 or 2
Default = 0
0=Disable, 1=Enable on LCD &
AO, 2=Enable on LCD, but not
AO for
dew point temperature
calculation and display.
Dew Point Method
0, 1, 2 or 3
Default = 0
Sets method used to calculate
dew point temperature.
Pipeline Pressure
0.0 to 500000
Default = 1000
Sets pressure at which dew
point temperature is calculated.
AI Pressure Input
0 or 1
Default = 0
Enables or disables usage of
pipeline pressure via analog
input.
AI 4 mA Value
0-500000
Default = 0.0
Sets value corresponding to 4
mA current loop input (mb).
AI 20 mA Value
0-500000
Default = 100000
Sets value corresponding to 20
mA current loop input (mb).
Modbus Address
User Set, 0-250
Default: 0
Sets the address for the
analyzer.
Modbus Mode
User Set, 0, 1, or 2
Default: 0
Sets type of Modbus protocol.
2-Way Com Port
0, 1, 2 or 3
Default = 1
Sets the communications port
for the port for which two-way
communications is allowed.
Baud Rate
0, 1, 2, 3 or 4
Default: 3
Sets RS-232 port baud rate.
Validation Wait Time
60-32000
Default = 60
The amount of time (in
seconds) for validation gas to
purge the system.
Val Attempts
1 to 8000
Default = 1
Sets the number of attempt(s)
to run during the perm tube
Validation Mode.
Val 1 Concentration
0 to Full Scale
Default: 100
Sets concentration of validation
gas supply #1.
Val 2 Concentration
0 to Full Scale
Default: 60
Sets concentration of validation
gas supply #2.
Summary of Contents for SpectraSensors SS2000e
Page 2: ......
Page 4: ......
Page 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 113: ......
Page 114: ...www spectrasensors com contact 4900002262 Rev B...