Function group
SENSOR DATA
Sensor data, such as nominal diameter, calibration factor, etc., are set in the factory. All characteristic
values of the sensor are stored in the DAT memory.
Caution!
Normally these characteristic sensor data may not be altered. A change to the sensor data affects a
number of functions of the whole measuring system, especially its accuracy. The following functions of
this group can therefore only be changed after entering a
service code
and cannot be altered using
the personal code.
Please contact your E+H Service organisation for more information.
K-FACTOR
POS.
Display of the actual calibration factor of the sensor for positive flow direction.
The calibration factor is determined and set in the factory:
6
5-digit number with fixed decimal point: 0.5000...2.0000
Factory setting:
dependent
on nominal diameter and calibration
Caution!
Under normal operating conditions the calibration factor is not to be changed.
The service code number required for this is known to your E+H Service
organisation. Please contact E+H Service for any problems or questions that
have occurred.
K-FACTOR
NEG.
Display of the actual calibration factor of the sensor for negative flow direction.
The calibration factor is determined and set in the factory:
6
5-digit number with fixed decimal point: 0.5000...2.0000
Factory setting:
dependent
on nominal diameter and calibration
Caution!
Under normal operating conditions the calibration factor is not to be changed.
The service code number required for this is known to your E+H Service
organisation. Please contact E+H Service for any problems or questions that
have occurred.
ZERO POINT
Display of the zero point correction of the sensor. This value is determined
and set in the factory:
6
max. 4-digit number: –1000...+1000
Factory setting:
dependent
on nominal diameter and calibration
Caution!
Under normal circumstances the zero point correction is not to be altered.
The service code number required for this is known to your E+H Service
organisation. Please contact E+H Service for any problems or questions that
have occurred.
NOMINAL
DIAMETER
Display of the actual nominal diameter of the sensor.
Caution!
The nominal diameter given may, in general, not be altered. Numerous
functions depend directly on the nominal diameter (technical units,
full-scale values, switch points, creepage, etc.). When the nominal diameter
is changed, all dependent parameters are set to a
new
plausible value!
6
Value between 2...2000 mm or 1/12... 78“
Factory setting:
dependent
on the sensor diameter
7
UNIT ==> NOM. DIAM. UNIT
Unit selection
→
see function “NOM. DIAM. UNIT”
Caution!
Caution!
Caution!
Caution!
Caution!
Promag 33
6 Description of Functions
Hauser
81
Summary of Contents for PROMAG 33
Page 43: ...Promag 33 5 Display and Operation Endress Hauser 43...
Page 46: ...5 Display and Operation Promag 33 46 Endress Hauser...
Page 84: ...6 Description of Functions Promag 33 84 Endress Hauser...
Page 98: ...7 Trouble shooting Maintenance and Repairs Promag 33 98 Endress Hauser...
Page 106: ...8 Dimensions Promag 33 106 Endress Hauser...
Page 116: ...9 Technical Data Promag 33 116 Endress Hauser...
Page 125: ......