Function group
SYSTEM PARAMETERS
ACCESS CODE
All data of the Promag 33 measuring system are protected against
unauthorised access. Only by first entering a code number in this function
programming is enabled and the settings of the instrument can then be
altered. If, in any function, the
6
operating elements are touched, then the
measuring system jumps automatically into this function and the display shows
the prompt to enter the code number (if programming is locked):
→
Enter code number 33 (factory setting) or
→
Enter personal code number (see “DEF. PRIVATE CODE”, page 77)
Note!
•
After jumping to the HOME position programming is again locked after
60 seconds if no operating element is touched during this time.
•
Programming can also be locked by entering any number (not the customer
code number) in this function.
•
If you can no longer find your personal code number, then the
Hauser service organisation will be pleased to help you.
•
Certain functions can only be modified by entering a special service code
number. This is known to your E+H Service organisation. Please contact
your E+H Service for any further information you may require.
6
max. 4-digit number: 0...9999
Factory setting:
0
SELF
CHECKING
Switching the periodical self check of the amplifier on or off.
The amplifier is fitted with an automatic temperature compensation.
Any temperature drift occurring in the region of the amplifier path can be
compensated for by a periodical measurement against an internal reference
voltage.
Note!
This function is not available if the “BATCH VARIABLE” (see page 64) is set to
“VOLUME”. In this case no periodic self check is carried out.
6
OFF –
ON
Note!
Note!
6 Description of Functions
Promag 33
78
Hauser
Summary of Contents for PROMAG 33
Page 43: ...Promag 33 5 Display and Operation Endress Hauser 43...
Page 46: ...5 Display and Operation Promag 33 46 Endress Hauser...
Page 84: ...6 Description of Functions Promag 33 84 Endress Hauser...
Page 98: ...7 Trouble shooting Maintenance and Repairs Promag 33 98 Endress Hauser...
Page 106: ...8 Dimensions Promag 33 106 Endress Hauser...
Page 116: ...9 Technical Data Promag 33 116 Endress Hauser...
Page 125: ......