Connection diagram for the transmitter (“HART”)
3
28
20
21
23
24
25
26
27
1
2
22
+
-
+
-
+
-
+
-
1
2
3
28
20
21
22
23
24
25
26
27
Supply cable
Ground terminal for
protective conductor
Ground terminal for
signal cable shield
Signal cable
ba009
y
3
5
Fuse
•
Power supply 20...55 V AC / 16...62 V DC:
2.5 A slow-blow / 250 V; 5.2
×
20 mm
•
Power supply 85...260 V AC:
1 A slow-blow / 250 V; 5.2
×
20 mm
Ground connection (protective earth)
L1
for AC power supply
L+
for DC power supply
N
L–
Pulse/frequency output (active/passive): f
max
= 10 kHz
active: 24 V DC, 25 mA (max. 250 mA/20 ms)
passive: 30 V DC, 250 mA
Alarm output (Relay 1): can be configured
max. 60 V AC / 0.5 A AC; max. 30 V DC / 0.1 A DC
Current output (active):
0/4...20 mA, R
L
< 700
Ω
(with HART: R
L
≥
250
Ω
)
Ground connection (screening of signal cable)
The input and outputs are galvanically isolated from the power supply and
from each other.
Version with a “HART” communication module
Status output (Relay 2): can be configured
max. 60 V AC / 0.5 A AC; max. 30 V DC / 0.1 A DC
Fig. 22
Terminal compartment
Promag 33 (HART)
Promag 33
4 Electrical Connection
Hauser
23
Summary of Contents for PROMAG 33
Page 43: ...Promag 33 5 Display and Operation Endress Hauser 43...
Page 46: ...5 Display and Operation Promag 33 46 Endress Hauser...
Page 84: ...6 Description of Functions Promag 33 84 Endress Hauser...
Page 98: ...7 Trouble shooting Maintenance and Repairs Promag 33 98 Endress Hauser...
Page 106: ...8 Dimensions Promag 33 106 Endress Hauser...
Page 116: ...9 Technical Data Promag 33 116 Endress Hauser...
Page 125: ......