Function group
COMMUNICATION
START PULSE
WIDTH
Certain functions of the auxiliary input are only started via a pulsed voltage
(see page 70). In this function, you enter the minimum pulse width to be
reached by the input pulse in order that the appropriate function is activated.
This ensures that the function is not activated by transient voltage peaks
(interference pulses).
Note!
This function is only available if the Promag electronics are fitted with the
communications module “RS 485” and if the auxiliary input is activated as
well as appropriately configured.
6
max. 3-digit number: 20...100 ms
Factory setting:
20 ms
SYSTEM
CONFIG.
In this function, the actual configuration of the communications module
“RS 485” is shown:
Note!
•
This function is only available, if the transmitter electronics are fitted with a
RS 485 communication module.
•
This function can only be reconfigured after entering a special service code.
If you have problems with the existing configuration, then please contact
your E+H Service organisation.
Display:
RS 485 / CURRENT
1) 2)
RS 485 / FREQUENCY
2) 3)
AUX. INPUT / CURRENT
1) 4)
AUX. INPUT / FREQ.
3) 4)
1)
The function group “PULSE / FREQ. OUTPUT” is blanked out.
2)
The matrix fields for auxiliary input are blanked out.
3)
The function group “CURRENT OUTPUT” is blanked out.
4)
The matrix fields for RS 485 are blanked out.
Note!
Note!
Promag 33
6 Description of Functions
Hauser
71
Summary of Contents for PROMAG 33
Page 43: ...Promag 33 5 Display and Operation Endress Hauser 43...
Page 46: ...5 Display and Operation Promag 33 46 Endress Hauser...
Page 84: ...6 Description of Functions Promag 33 84 Endress Hauser...
Page 98: ...7 Trouble shooting Maintenance and Repairs Promag 33 98 Endress Hauser...
Page 106: ...8 Dimensions Promag 33 106 Endress Hauser...
Page 116: ...9 Technical Data Promag 33 116 Endress Hauser...
Page 125: ......