27
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0)
255 (10.0)
134 (5.3)
30 (1.18)
A0029553
9
Weather protection cover for Proline 500
Cover locking: Proline 500
NOTICE
Order code for "Housing", option L "Cast, stainless": The covers of the transmitter
housing are provided with a borehole to lock the cover.
The cover can be locked using screws and a chain or cable provided by the customer.
‣
It is recommended to use stainless steel cables or chains.
‣
If a protective coating is applied, it is recommended to use a heat shrink tube to protect
the housing paint.
1
1
2
2
1
1
5 (0.2) min. 15 (0.6)
SW 2.5
M3
A0029799
1
Cover borehole for the securing screw
2
Securing screw to lock the cover
Determining the pitch angle and roll angle
For correct measurement, the pitch angle and roll angle must be determined and
entered in the
Installation angle pitch
Installation
angle roll
The angle can be entered with an accuracy of ±10 °.