Proline Promass Q 500 PROFIBUS PA
Installation
Hauser
29
2. Install the measuring device or turn the transmitter housing so that the cable entries
do not point upwards.
A0029263
6.2.4
Mounting the transmitter housing: Proline 500 – digital
L
CAUTION
Ambient temperature too high!
Danger of electronics overheating and housing deformation.
‣
Do not exceed the permitted maximum ambient temperature .
‣
If operating outdoors: Avoid direct sunlight and exposure to weathering, particularly in
warm climatic regions.
L
CAUTION
Excessive force can damage the housing!
‣
Avoid excessive mechanical stress.
The transmitter can be mounted in the following ways:
• Post mounting
• Wall mounting
Post mounting
L
WARNING
Excessive tightening torque applied to the fixing screws!
Risk of damaging the plastic transmitter.
‣
Tighten the fixing screws as per the tightening torque: 2 Nm (1.5 lbf ft).
ø
20…70
( 0.79…2.75)
ø
~102 (~ 4.0)
4 x
SW 10
3 x
TX 25
A0029051
10
Engineering unit mm (in)