Technical data
Proline Promass Q 500 PROFIBUS PA
258
Hauser
16.8 Environment
Ambient temperature
range
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
Storage temperature
–50 to +80 °C (–58 to +176 °F)
Climate class
DIN EN 60068-2-38 (test Z/AD)
Degree of protection
Transmitter
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
Sensor
As standard: IP66/67, type 4X enclosure
External WLAN antenna
IP67
Vibration resistance
• Vibration, sinusoidal according to IEC 60068-2-6
– 2 to 8.4 Hz, 3.5 mm peak
– 8.4 to 2
000 Hz, 1 g peak
• Vibration broad-band random, according to IEC 60068-2-64
– 10 to 200 Hz, 0.003 g
2
/Hz
– 200 to 2
000 Hz, 0.001 g
2
/Hz
– Total: 1.54 g rms
Shock resistance
Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Impact resistance
Rough handling shocks according to IEC 60068-2-31
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
For details, refer to the Declaration of Conformity.
16.9 Process
Medium temperature range
Standard
–50 to +205 °C (–58 to +401 °F)
Low temperature
–196 to +150 °C (–320 to +302 °F) Order code for "Measuring tube
material", option
LA