Functions relevant to custody transfer
Commissioning Instructions for PTB gas approval Proline Promass 84
38
Hauser
7.27
Basic function
Process parameter
Adjustment
7.28
Basic function
Process para.
Pressure correct.
7.29
Basic function
System parameter
Configuration
7.30
Basic function
Sensor data
Configuration
Function name
No.
Description
Configuration
ZERO POINT ADJUST
6480 Use this function to activate zero point adjustment.
Options:
e.g. CANCEL
DENSITY ADJUST
MODE
6482 Use this function to select whether a 1-point or 2-point
density adjustment is to be carried out.
Options:
e.g. CANCEL
DENSITY SETPOINT 1
6483 Use this function to enter the density setpoint value for
the 1st fluid for which a density adjustment is to be
carried out.
User input:
e.g. 1.0000 kg/l
MEASURE FLUID 1
6484 Use this function to measure the current density of the
first fluid for density adjustment.
User input:
e.g. CANCEL
DENSITY SETPOINT 2
6485 Use this function to enter the density setpoint value for
the 2nd fluid for which a density adjustment is to be
carried out.
User input:
e.g. 1.0000 kg/l
MEASURE FLUID 2
6486 Use this function to enter the current density of the
second fluid for density adjustment.
User input:
e.g. CANCEL
DENSITY ADJUST
6487 Use this function to activate the density adjustment or
measurement of the first or second fluid.
User input:
e.g. CANCEL
RESTORE ORIGINAL
6488 Use this function to restore the original density
coefficients determined at the factory.
User input:
e.g. NO
Function name
No.
Description
Configuration
PRESSURE MODE
6500 Use this function to configure the automatic pressure
correction.
Options: e.g. OFF
PRESSURE
6501 Use this function to enter the pressure value for
pressure correction.
User input: e.g. 0 barg
Function name
No.
Description
Configuration
INSTALLATION DIRECTION
SENSOR
6600 Use this function to change the sign for the flow
variable.
Options:
e.g. NORMAL
DENSITY DAMPING
6602 Use this function to configure the sensitivity of the
density measuring signal to fluctuations in fluid density.
User input:
e.g. 0.00 s
FLOW DAMPING
6603 Use this function to configure the sensitivity of the
measuring signal to interference peaks.
User input:
e.g. 0 s
POSITIVE ZERO RETURN
6605 Use this function to interrupt the evaluation of the
measured variables.
Options:
e.g. OFF
Function name
No.
Description
Configuration
K-FACTOR
6800 Use this function to display the current calibration factor
for the sensor.
Display: depending on
DN and calibration
ZERO POINT
6803 Use this function to display the current zero point
correction value for the sensor.
Display: depending on
calibration
NOMINAL DIAMETER
6804 Use this function to display the nominal diameter of the
sensor.
Display: depending on
DN
POSITIVE ZERO RETURN
6605 Use this function to interrupt the evaluation of the
measured variables.
Options:
e.g. OFF