Functions relevant to custody transfer
Commissioning Instructions for PTB gas approval Proline Promass 84
36
Hauser
7.20
Outputs
Relay output
Configuration
7.21
Outputs
Relay output
Operation
7.22
Inputs
Status input
Configuration
7.23
Inputs
Status input
Operation
Function name
No.
Description
Configuration
ASSIGN RELAY
4700 Use this function to assign a switching function to the
relay output.
Options:
e.g. FAULT MESSAGE
ON-VALUE
4701 Use this function to enter the value for the switch-on
point.
User input:
e.g. 0 kg
SWITCH-ON DELAY
4702 Time period which must elapse after the switch-on
point has been reached and before the output switches.
User input:
e.g. 0.0 s
OFF-VALUE
4703 Use this function to enter the value for the switch-off
point.
User input:
e.g. 2 kg
SWITCH-OFF DELAY
4704 Time period which must elapse after the switch-off
point has been reached and before the output switches.
User input:
e.g. 0.0 s
MEASURING MODE
4705 Use this function to select the measuring mode.
Options:
e.g. STANDARD
TIME CONSTANT
4706 Time constant which determines how the relay output
reacts to severely fluctuating flow variables.
User input:
e.g. 0.00 s
Function name
No.
Description
Configuration
ACTUAL STATUS RELAY
4740 Use this function display the status currently output.
Display:
e.g. MAKE CONTACT
OPEN
SIMULATION SWITCH
POINT
4741 Use this function to activate the simulation of the relay
output.
Options:
e.g. OFF
VALUE SIMULATION
SWITCH POINT
4742 Default value for the simulation of the relay output.
User input:
e.g. MAKE CONTACT
OPEN
Function name
No.
Description
Configuration
ASSIGN STATUS INPUT
5000 Use this function to assign a switching function to the
status input.
Options: e.g. RESET
FAULT MESSAGE
ACTIVE LEVEL
5001 Use this function to select the level at which the
switching function is to be activated.
User input:
e.g. HIGH
MINIMUM PULSE WIDTH
5002 Use this function to specify the minimum pulse width
which must be present for the switching function to be
activated.
User input:
e.g. 50 ms
Function name
No.
Description
Configuration
SIMULATION STATUS
INPUT
5041 Use this function to activate the simulation of the status
input.
Options:
e.g. OFF
VALUE SIMULATION
STATUS INPUT
5042 Default value for the simulation of the status input.
User input:
e.g. LOW