Technical data

Commissioning Instructions for PTB gas approval Proline Promass 84

26

Hauser

6.1.8

Normal operating conditions

• Measured variable: mass of gas flowing through in kg, volume of gas flowing through in m

3

 

(possible with clean gases)

• Gas types, covered by MI-002 : fuel gases, technically pure flammable gases and mixtures of these 

gases in a gaseous state

• Additional allowed gas types: Nonflammable technical gases, mixtures of these and mixtures 

containing fuel gases

• Gas temperature range: –25°C to +55°C
• Operating overpressure range:

– 5 to 100 bar (standard)
– to 280 bar (Promass O)
– to 350 bar (high-pressure version Promass A)

• Climatic ambient conditions: –40°C to +55°C; condensation possible
• Degree of protection: IP67
• Mechanical ambient conditions: M3 (for transporting on vehicles), M2 (in operation)
• Electromagnetic ambient conditions: E3

6.1.9

Human interface

Display elements

• Liquid crystal display: illuminated, four lines with 16 characters per line
• Selectable display of different measured values and status variables
• 3 totalizers
• At ambient temperatures below –20 °C, the readability of the display may be impaired.

Operating elements

• Local operation with three optical sensors (

S

/

O

/

F

)

• Application-specific quick setup menus for straightforward commissioning

Remote operation

Depending on device version:
• Operation by means of HART protocol
• Operation by means of Modbus RS485

Summary of Contents for Proline Promass 84

Page 1: ...s of version HART V 3 01 XX device software Modbus V 3 06 XX device software Commissioning Instructions for PTB gas approval DE 08 MI002 PTB007 Proline Promass 84 Coriolis gas meter for applications s...

Page 2: ......

Page 3: ...ay Control Unlocking locking 30 7 8 User interface Main line Configuration 30 7 9 User interface Main line Multiplex 30 7 10 User interface Additional line Configuration 30 7 11 User interface Additio...

Page 4: ...ts the following documentation for the Proline Promass 84 HART measuring device BA00109D Operating Instructions BA00110D Description of Device Functions for the Proline Promass 84 Modbus RS485 measuri...

Page 5: ...llowing The chapters Additional documentation 4 Der W M Device Viewer Enter the serial number from the nameplate www endress com deviceviewer Reorder The measuring device is reordered using the order...

Page 6: ...ters and Volume Conversion Devices and also has an EC type examination certificate DE 08 MI002 PTB007 Two versions are available Compact version Transmitter and sensor form a single mechanical unit Re...

Page 7: ...nameplate 1 Certificate recipient 2 Name of the transmitter 3 Order code meaning of individual letters and digits 11 4 Serial number Ser no 5 Extended order code Ext ord cd 6 Year of manufacture 7 Po...

Page 8: ...class 4 Permitted range of gas flow 5 Permitted pressure range 6 Permitted gas temperature range 7 Permitted operating temperature range 8 Information on medium 9 Information on pulse value 10 Confor...

Page 9: ...er code Ext ord cd 6 Year of manufacture 7 Ambient temperature range 8 Degree of protection 9 Calibration factor with zero point K factor 10 Max fluid temperature Tm 11 Accuracy of density measurement...

Page 10: ...nd outputs 8 Version of device software currently installed Device SW 9 Installed communication type Communication 10 Information on current communication software Drivers Device Revision and Device D...

Page 11: ...sensor and is structured as follows P R O M A S S 8 4 Pos no 1 Instrument family 2 Electronics 3 Sensor A F O or X 4 to 5 Nominal diameter 6 Hyphen 7 Measuring tube material 8 to 10 Process connectio...

Page 12: ...Terminal no inputs outputs 20 21 22 23 24 25 26 27 Fixed communication boards permanent assignment S Frequency output Ex i passive Current output HART Ex i active T Frequency output Ex i passive Curre...

Page 13: ...55 V AC 16 to 62 V DC Terminal no 1 L1 for AC L for DC Terminal no 2 N for AC L for DC b Signal cable Terminal assignment 12 c Fieldbus cable Terminal assignment 12 d Ground terminal for protective gr...

Page 14: ...9 10 pink 11 12 white 41 42 brown 3 3 1 Cable specification connecting cable In the remote version the specifications of the cable connecting the transmitter to the sensor are as follows 6 0 38 mm PV...

Page 15: ...et up depending on the measuring device version Hardware write protection on off Device address Terminating resistors Configuration of current output active passive Configuration of relay output NC co...

Page 16: ...ected The types serial numbers and software versions of all the meter modules used at the point of installation must be compared with the specifications in the accompanying documentation Likewise the...

Page 17: ...must be technically ensured by the operator of the meters The functioning of the uninterruptible voltage supply and the triggering of the automatic maintenance request must be checked and recorded in...

Page 18: ...Operate the system until operating conditions have settled This usually takes approx 5 minutes at maximum flow rates 2 Stop the flow v 0 m s 3 Check the shutoff valves for leaks 4 Check that the opera...

Page 19: ...as output configuration custody transfer variable and measuring mode Note A detailed description of the functions can be found in the manual entitled Description of Device Functions BA00110D Promass 8...

Page 20: ...7 Sealing of the measuring device a0001778 Fig 7 Examples of how to seal the various device versions Nicht unter Spannung ffnen K e e p c o v e r t i g h t w h i l e c i r c u i t s a r e a l i v e N...

Page 21: ...nt version can be found in the Operating Instructions BA00109D Promass 84 HART or BA00129D Promass 84 Modbus RS485 4 Remove the S DAT 5 Reconnect the device to the power supply 6 The device runs throu...

Page 22: ...nosis of intelligent field devices The use of status information means that you also have a simple but effective tool for monitoring devices Proline flowmeters are accessed via a service interface or...

Page 23: ...erpressure Pe min DN Minimum flow Qmin kg h Maximum flow Qmax kg h Lowest permitted density at pe min kg m3 mm inch 8 3 8 3 6 20 pe min 3 6 15 12 0 60 pe min 4 0 25 1 30 0 200 pe min 3 0 40 1 135 0 40...

Page 24: ...passive selectable galvanically isolated time constant selectable 0 05 to 100 s full scale value selectable temperature coefficient typically 0 005 o r C resolution 0 5 A Active 0 4 to 20 mA RL 700 P...

Page 25: ...number is output for the process variables Switch output Relay output Normally closed NC or break or normally open NO or make contacts available factory setting normally open max 30 V 0 5 A AC 60 V 0...

Page 26: ...O to 350 bar high pressure version Promass A Climatic ambient conditions 40 C to 55 C condensation possible Degree of protection IP67 Mechanical ambient conditions M3 for transporting on vehicles M2 i...

Page 27: ...ass A Pressure measuring device approval Measuring devices with a nominal diameter smaller than or equal to DN 25 correspond to Article 3 3 of the EC Directive 97 23 EC Pressure Equipment Directive an...

Page 28: ...f relevant to custody transfer CURRENT OUTPUT 1 CUSTODY TRANSFER Z003 Use this function to select current output 1 for transferring the CT signal Options YES if relevant to custody transfer TOTALIZER...

Page 29: ...g CENT FACTOR ARBITRARY MASS 0601 Use this function to enter quantity factor without time for arbitrary mass mass flow User input e g 1 TEXT ARBITRARY VOLUME 0602 Use this function to enter text for a...

Page 30: ...SSIGN 2200 Use this function to assign a display value to the main line Options e g MASS FLOW 100 VALUE 2201 Use this function to specify the value to be displayed as 100 User input 5 digit floating p...

Page 31: ...value to be displayed as 100 User input 5 digit floating point number FORMAT 2602 Use this function to select the number of decimal places Options e g XX XXX DISPLAY MODE 2603 Use this function to sel...

Page 32: ...ion to select the current span Options e g 4 20 mA VALUE 0_4 mA 4002 Use this function to specify the value at which a current value of 0 or 4 mA is output Options e g 0 kg h VALUE 20 mA 4003 Use this...

Page 33: ...o select the measuring mode Options e g STANDARD OUTPUT SIGNAL 4207 Use this function to select the output configuration Options e g PASSIVE POSITIVE TIME CONSTANT 4208 Time constant which determines...

Page 34: ...LAY 4243 Time period which must elapse after the switch on point has been reached and before the output switches User input e g 0 0 s OFF VALUE 4244 Use this function to enter the value for the switch...

Page 35: ...EQUENCY 4303 Default value for the simulation of the frequency output User input e g 0 Hz Function name No Description Configuration SIMULATION PULSE 4322 Use this function to activate the simulation...

Page 36: ...Time constant which determines how the relay output reacts to severely fluctuating flow variables User input e g 0 00 s Function name No Description Configuration ACTUAL STATUS RELAY 4740 Use this fun...

Page 37: ...ALUE HIGH 6424 Use this function to enter the upper density value for fluid monitoring User input e g 6 0000 kg l EPD RESPONSE TIME 6425 Use this function to specify the time period which must elapse...

Page 38: ...e first or second fluid User input e g CANCEL RESTORE ORIGINAL 6488 Use this function to restore the original density coefficients determined at the factory User input e g NO Function name No Descript...

Page 39: ...unction to display the density coefficient C 2 display only DENSITY COEFFICIENT C 3 6853 Use this function to display the density coefficient C 3 display only DENSITY COEFFICIENT C 4 6854 Use this fun...

Page 40: ...the measuring system Options NO Function name No Description Configuration DEVICE SOFTWARE 8100 Use this function to display the current version of the device software display only Function name No D...

Page 41: ...rent in mA display only CARRIER TUBE TEMPERATURE 9002 Use this function to display the current temperature of the carrier tube display only Function name No Description Configuration MONITORING MODE 9...

Page 42: ...g point number incl sign Factory setting depending on the option selected in the gas type function PRESSURE MODE 9204 Use this function to select the setting for pressure measuring mode within ethylen...

Page 43: ...Commissioning Instructions for PTB gas approval Proline Promass 84 Functions relevant to custody transfer Endress Hauser 43...

Page 44: ...www endress com worldwide SD00128D 06 EN 13 11 71144260 FM SGML9 0 ProMoDo...

Reviews: