Commissioning Instructions for PTB gas approval Proline Promass 84
Functions relevant to custody transfer
Hauser
41
7.38
Supervision
Version-info
I/O module
7.39
Service & Analysis
Application
Test points
7.40
Service & Analysis
Application
Monitoring
Function name
No.
Description
Configuration
I/O MODULE TYPE
8300 Use this function to display the configuration of the I/O
module complete with terminal number.
display only
SOFTWARE REV. NUMBER
I/O MODULE
9303 Use this function to display the software revision
number of the I/O module.
display only
Function name
No.
Description
Configuration
OSCILLATION FREQUENCY
9000 Use this function to display the current operating
frequency (resonance frequency) of the measuring tubes
in Hz. This depends on the sensor type and fluid
density.
display only
EXCITATION CURRENT
9001 Use this function to display the effective value of the
excitation current in mA.
display only
CARRIER TUBE
TEMPERATURE
9002 Use this function to display the current temperature of
the carrier tube.
display only
Function name
No.
Description
Configuration
MONITORING MODE
9100 Use this function to suppress / output the following
error messages:
• #701 EXC. CURR. LIM.
• #701 FLUID INHOM.
• #701 NOISE LIM. CH0
• #701 NOISE LIM. CH0
• #701 FLOW LIMIT
Options:
OFF = Error messages
are suppressed.
ON = Error messages
are output.
OSCILLATION AMPLITUDE
9101 Use this function to increase or decrease the oscillating
amplitude of the measuring tubes relative to the
standard value of 100%.
User input: 50 to 150%
of standard value
GAIN AMPLIFIER
9102 Use this function to enter the relative signal
amplification
User input: 25 to 150%
of standard value
MONITOR INTERVAL
9102 Use this function to specify the time from when a fault
occurs to when it is signaled by means of an error
message on the onsite display or at the outputs.
This interval also applies to the time from when the
fault is eliminated to when the error message is
removed from the onsite display or the outputs.
User input:
1.0 to 100 s