Commissioning
Proline Promag 55 PROFIBUS PA
70
Hauser
!
Note!
• The display returns to the cell SETUP COMMISSIONING (1002) if you press the
X
key combination during
parameter interrogation. The stored parameters remain valid.
• The "Commissioning" Quick Setup must be carried out before any of the other Quick Setups described in these
Operating Instructions are run.
• The system units selected via the Quick Setup only apply for displaying on the local display. They do not affect the
measured variables (Volume Flow, Mass Flow) that are transmitted via PROFIBUS.
m
The "DELIVERY SETTINGS" option sets every selected unit to the factory setting.
The "ACTUAL SETTINGS" accepts the units you configured beforehand.
n
Only units not yet configured in the current setup are offered for selection in each cycle. The unit for mass and
volume is derived from the corresponding flow unit.
o
The "YES" option remains visible until all the units have been configured.
"NO" is the only option displayed when no further units are available.
p
The "automatic parameterization of the display" option contains the following basic settings/factory settings:
YES
Main line = Volume Flow
Additional line = Totalizer 1
Information line = Operating/System Conditions
NO
The existing (selected) settings remain.