Installation
Proline Promag 55 PROFIBUS PA
36
Hauser
Installation in control panel
1.
Prepare the opening in the panel as illustrated.
2.
Slide the housing into the opening in the panel from the front.
3.
Screw the fasteners onto the wall-mount housing.
4.
Place the threaded rods in the fasteners and screw them down until
the housing is seated tightly against the panel. Afterwards, tighten the locking nuts.
Additional support is not necessary.
a0001131-ae
Fig. 28:
Panel Installation (wall-mount housing)
Pipe mounting
The assembly should be performed by following the instructions in the following diagram.
"
Caution!
If the device is mounted to a warm pipe, make sure that the housing temperature does not exceed
+60 °C (+140 °F), which is the the maximum permissible temperature.
a0001132-ae
Fig. 29:
Pipe mounting (wall-mount housing)
245 (9.65)
~110 (~4.33)
210 (8.27)
+0.5 (+0.019)
–0.5 (–0.019)
+0.5 (+0.019)
–0.5 (–0.019)
mm (inch)
Ø 20…70
(Ø 0.79…2.75)
~
~ 6.1)
155 (
mm (inch)