Troubleshooting
Proline Promag 55 PROFIBUS PA
106
Hauser
No. # 1xx
→
Software error
101
S: GAIN ERROR AMP.
$
: # 101
0x0F
BAD
Device Failure
Constant
Gain Error
Amplifier
Cause:
Gain deviation compared to reference gain is
greater than 2%.
Remedy:
Replace the amplifier board
121
S: A/C COMPATIB.
!: # 121
0x0F
BAD
Device Failure
Constant
Amplifier and I/O
board only partially
compatible
Cause:
Due to different software versions, I/O board and
amplifier board are only partially compatible
(possibly restricted functionality).
!
Note!
– The indication on the display as notice message
appears only for 30 seconds (with listing in
"Previous system condition" function).
– This condition can occur if only one electronics
board has been exchanged; the extended
software functionality is not available. The
previously existing software functionality is still
working and the measurement possible.
Remedy:
Module with lower software version must either be
updated by FieldCare using the required
(recommended) software version, or the module
must be replaced
No. # 2xx
→
Error in DAT / no communication
205
S: LOAD T-DAT
!: # 205
0x0F
BAD
Device Failure
Constant
Save to T-DAT
failed
Cause:
Data backup (downloading) to T–DAT failed, or
error when accessing (uploading) the calibration
values stored in the T–DAT.
Remedy:
1.
Check whether the T-DAT is correctly plugged
into the amplifier board.
2.
Replace the T-DAT if it is defective.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics.
Check the:
– Spare part set number
– Hardware revision code
3.
Replace measuring electronics boards if
necessary
206
S: SAVE T-DAT
!: # 206
0x0F
BAD
Device Failure
Constant
Restore from T-
DAT failed
261
S: COMMUNICAT. I/O
$
: # 261
0x18
0x19
0x1A
BAD
No
Communicati
on
O.K.
Low
High
Communication
failure
Cause:
Communication error. No data reception between
amplifier and I/O board or faulty internal data
transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
Q
u
a
lit
y
c
o
d
e
(
H
EX
)
Me
as
ure
d
val
u
e
s
ta
tus
Qu
a
lit
y
s
ta
tu
s
Qu
a
lit
y
s
u
b
st
a
tu
s
Li
m
its