Troubleshooting
Proline Promag 55 PROFIBUS PA
108
Hauser
692
S: SIM. MEASURAND
!: # 692
0x60
0x61
0x62
UNCERTAIN
Simulated
Value
(manually
specified
value)
O.K.
Low
High
Simulation
measured value
active
Cause:
Simulation of the measured value is active.
Remedy:
Switch off simulation.
Access:
SUPERVISION
→
SYSTEM
→
OPERATION
→
SIM. MEASURAND (
→
OFF)
698
S: DEV. TEST ACT.
!: # 698
0x60
0x61
0x62
UNCERTAIN
Simulated
Value
(manually
specified
value)
O.K.
Low
High
Device test via
Fieldcheck active
Cause:
The measuring device is being checked on site via
the test and simulation device.
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy